Do you know that clearing is not just about light sheet microscopy? Even if you have done your job well and your sample is directly on the coverslip (not on the slide), as soon as your sample is thicker than 10 um (1 cell diameter), you will see the effect of the refraction index mismatch.
What is that? Your sample and the mounting medium around it have a certain refraction index (or likely several). The objective you are using is designed for a certain refraction index (e.g. air, water or oil). If these refraction indices do not match what happens? as soon as you image a tiny bit away from the coverslip, the sample will look elongated, the intensity and contrast will drop very fast.
Sounds familiar? If yes, changing your mounting medium to match the objective will solve the problem. It works for light sheet but it also works for wide field or confocal imaging! Just change your mounting medium and you will see an enormous difference!
Here is an article describing a one-step clearing protocol. This basically is about using a different mounting medium. Easy, cheap and non-toxic! Give it a try!
Have a look at this post for more info.